EasyManuals Logo

GE Datex-Ohmeda Aisys Carestation User Manual

GE Datex-Ohmeda Aisys Carestation
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #272 background imageLoading...
Page #272 background image
Aisys
7-90 04/08 M1046983
7.10.10 eVap Temperature troubleshooting
Flowchart
25
Is the Error entry
“Cass Temp”,
“Cass x-ducer”,
or “Manifold”
related?
Start
Error Log Entry
CONDENSATION CONDITIONS EXIST
VAP CONDENSATION CONDITION EXISTS
CASSETTE TEMPERATURE EEPROM FAILURE
CASSETTE TEMPERATURE FAILURE
MAN CASS OVER UNDER TEMP
MANIFOLD TEMPERATURE EEPROM FAILURE
MANIFOLD TEMPERATURE FAILURE
Cassette Temperature Related
CONDENSATION CONDITIONS EXIST
VAP_CONDENSATION_CONDITION_EXISTS
Cassette x-ducer Related
CASSETTE TEMPERATURE FAILURE
CASSETTE TEMPERATURE EEPROM FAILURE
MANIFOLD TEMPERATURE EEPROM FAILURE
Manifold Temperature Related
MANIFOLD TEMPERATURE FAILURE
MAN CASS OVER UNDER TEMP
Cass Temp
Cass x-ducer
Manifold
Indicates a warm
cassette was installed.
No service action required.
Replace
Cassette Temperature Sensor.
•the ADB Board
Check connection of the temperature
sensors to the ADB.
If the problem persists,
check the value of the temp sensors using
the PC Service App.
If reading does not change when the temp
cable is disconnected,
replace the ADB.
If the reading does change,
replace the temp sensor out of range.
VAP CASS TEMP SENSOR COMPAT FAIL
(Section 12.10.1)
AB.75.182

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters