Aisys
7-86 04/08 M1046983
7.10.6 eVap Pressure troubleshooting
Flowchart
21
Subassembly.
Is the difference between the
Mixer Outlet and Cassette
within -0.8 and 1.4 PSI?
Error Log Entry
CASSETTE PRESSURE ERROR
FLOW MANIFOLD EEPROM FAILURE
CASSETTE PRESSURE ERROR
Start
Within the PC Service Application, with the
Inflow Valve and Scavenger Valve open,
establish 500 ml flow from the Mixer
Subtract Mixer Outlet pressure from Cassette Pressure.
.
Replace the
Flowmeter
No
Contact Technical
Support for advice.
Yes
VAP FLOW METER BLK COMPAT FAIL
AB.75.178
(Section 12.3.4) (Section 12.3.2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters

Related product manuals