4 Install/Service Menus
M1046983 04/08 4-19
4.5.1 Spiro Calibration The Spiro Calibration instructions appear when the focus is on Spiro
Calibration menu item.
Instructions
Spiro Calibration menu
To display TV data:
1. Connect a spirometry sensor to MGas.
2. Push the ComWheel to continue.
3. Select the correct spirometry sensor type (Adult or Pedi).
4. Use the test device to deliver a known TV through the sensor.
5. If necessary, repeat steps 1-4 for the other type of sensor.
Menu Item Values Comments
Sensor Type Adult or Pedi Select Sensor type: Adult (Dlite) or
Pedi (Pedilite).
Insp Gain
TVinsp
1000
_ _ _ ml
Exp Gain
TVexp
1000
_ _ _ ml

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters

Related product manuals