Aisys
9-16 04/08 M1046983
9.8 Servicing Aladin
2
cassettes
Note Before service or returning a cassette to the manufacturer, make sure that the cassette
is empty. All Aladin
2
cassettes must be emptied before shipping. Package and ship the
cassettes in a suitable container.
9.8.1 Emptying an
Aladin
2
cassette
1. Connect an empty bottle to the filling system and hold it tight.
2. Turn the Aladin
2
cassette so that agent flows into the bottle and wait until the
cassette is empty.
3. To get the maximum amount of agent out, rock the cassette from left to right and tip
it forward and back several times.
4. Remove the bottle before returning the cassette to the horizontal position.
AB.60.038
AB.60.045

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters

Related product manuals