EasyManuals Logo

GE Datex-Ohmeda Aisys Carestation User Manual

GE Datex-Ohmeda Aisys Carestation
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
Aisys
2-16 04/08 M1046983
2.7 Anesthesia Control board
2.7.1 Overview The Anesthesia Control board (A) uses a Motorola MCF5407 Coldfire microcontroller
with 4M Flash and 16M error correcting DRAM. The Anesthesia Control board includes
6 UARTs with a 64 byte FIFO and RS-422 communications to interface with the Display
Unit, an accessory port, and anesthesia delivery subsystems located in the pan
electronic enclosure. These include the Gas Mixer, Electronic Vaporizer, and the
Ventilator Interface board.
Figure 2-15 • Anesthesia Control board
A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters

Related product manuals