DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4 - 11
Top Console adjustment controls
The system Top Console can be freely moved in any horizontal direction (if the power is on or if the
mechanism is not locked in position prior to powering down the system). The vertical displacement of
the Top Console is motor driven. The control buttons are located between the handles.
Power ON the LOGIQ E10 before performing the following steps.
To move the Top Console if Power is OFF
To manually release the console’s XY-mechanism (frogleg) if the power is OFF, see: Section 6-4
"Operator Panel movement" on page 6-13.
CAUTION
To avoid injury or damage, make sure nothing is within the range of motion before moving the
Top Console. This includes both objects and people.
Table 4-4 Top Console adjustment controls: Lock (1), Up/Down (2)
Steps Corresponding Graphic
1.
CAUTION
To avoid injury or damage, make sure nothing is within the range of motion before moving
the Top Console. This includes both objects and people.
Lock/Release button (1): unparking and
sideways displacement of the console.
Up/Down button (2): raising/lowering the
console.
To LOCK the Top Console:
Press and hold down the Lock/Release button (1), while pushing the Top Console in the locked
position.
Release the button.
To RAISE/LOWER the Top Console:
Press and hold the Up/Down button (2) accordingly to raise or lower the Top Console.
Release the button when the Top Console is at the desired height.
To move the floating Top Console:
The unlocked Top Console can be easily moved by pressing and holding the Lock/Release
button (1) and moving the Top Console in any horizontal (XY) direction.
Release the button when the Top Console is at the desired direction.