DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 2 Site Preparations 2 - 11
2-3-3 Desirable features
In addition to the floor plan suggestions shown in Figure 2-1 and Figure 2-2, a nearby waiting room and
a receptacle for bio–hazardous waste, like used probe sheaths is suggested. Grid represents 30.5 x
30.5 cm (1 x 1 foot).
2-3-4 Minimal floor plan suggestions
Floor Plan Suggestion 4.27 x 5.18 m (14 x 17 foot) Key
Figure 2-1 Floor Plan Suggestion 4.27 x 5.18 m (14 x 17 foot)
Item Description Item Description
1.
Secretary or doctors desk
2.
Lockable cabinet ordered by GE for its software and
proprietary manuals
3.
Film viewer
4.
Counter top
5.
Counter top and sink with hot and cold water
6.
Overhead light dimmer – dual level lighting (bright/dim)
7.
Emergency oxygen
8.
Suction line
9.
LOGIQ E10
10.
Dedicated power outlet – circuit breaker protected and
easily accessible
11.
Network interface
12.
45,7 cm (1.5 ft) distance of LOGIQ E10 from wall or
objects
13.
Stool
14.
Footswitch
15.
Storage for linens and equipment
16.
Examination table – 193 x 61 cm (76 x 24 inches)
17.
Lavatory and dressing room
18.
Door – at least 76 cm (30 inches)
4
1
2
18
5
6
9
16
17
15
8
7
10
11
12
14
3
1
2
1
2
AutoAuto
BB
RR
LL
CommentComment
ClearClear
Body Pattern
Ellipse
Body Pattern
Ellipse
ZoomZoom
MM
XX
CFCF
ZZ
YY
PWPW
CW
Elasto
3D/4D
13