EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V1

GE LOGIQ V1 User Manual

GE LOGIQ V1
381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
Measurement and Analysis
LOGIQ V2/LOGIQ V1 – User Guide 2-55
Direction 5610736-100 Rev. 9
Intensity (Echo level) measurement
To make an echo level measurement:
1. Press Measure key to enter into measure mode, if
necessary.
2. Adjust Caliper Change in the Primary Menu to Intensity.
NOTE: The available values for Caliper Change menu can be set
in Utility -> Measure -> Advanced preset.
3. To position the caliper at the start point, move the Trackball.
4. To fix the trace start point, press Set. The caliper changes to
an active caliper.
5. To trace the measurement area, move the Trackball around
the anatomy. A dotted line shows the traced area.
6. To complete the measurement, press Set. The system
displays the echo level, as EL __ dB, in the Results Window.
NOTE: The echo level measurement is only available on a frozen
image, not on a B-paused image.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals