Best practices while scanning
NOTE:  Refer to the Optimizing the Image sections for a complete 
discussion of each control. 
Acoustic Output Default Levels
In order to assure that an exam does not start at a high output 
level, the LOGIQ V2/LOGIQ V1 initiates scanning at a reduced 
default output level. This reduced level is preset programmable 
and depends upon the exam category and probe selected. It 
takes effect when the system is powered on or Patient is 
selected.
To modify acoustic output, adjust the Power Output level.
Raise the Acoustic Output only after attempting image 
optimization with controls that have no effect on Acoustic 
Output, such as Gain and TGC. 
Be sure to have read and understood control explanations for 
each mode used before attempting to adjust the Acoustic 
Output control or any control that can effect Acoustic Output. 
Use the minimum necessary acoustic output to get the best 
diagnostic image or measurement during an examination. 
Begin the exam with the probe that provides an optimum focal 
depth and penetration.