Probes and Biopsy
5-6 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
 Rev. 9
Probe Safety
Handling precautions 
Electrical shock hazard 
Ultrasound probes are highly sensitive medical instruments 
that can easily be damaged by improper handling. Use care 
when handling and protect from damage when not in use. DO 
NOT use a damaged or defective probe. Failure to follow these 
precautions can result in serious injury and equipment 
damage.
The probe is driven with electrical energy that can injure the 
patient or user if live internal parts are contacted by conductive 
solution:
•  DO NOT immerse the probe into any liquid beyond the 
level indicated by the immersion level diagram. Refer to the 
immersion illustration in the Probe Cleaning Process 
section. Never immerse the probe connector or probe 
adaptors into any liquid.
•  DO NOT drop the probes or subject them to other types of 
mechanical shock or impact.  Degraded performance or 
damage such as cracks or chips in the housing may result.
•  Prior to each use, visually inspect the probe lens and case 
area for cracks, cuts, tears, and other signs of physical 
damage. DO NOT use a probe which appears to be 
damaged until you verify functional and safe performance. 
You must perform a more thorough inspection, including 
the cable, strain relief, and connector, each time you clean 
the probe. 
•  Before inserting the connector into the probe port, inspect 
the probe connector pins. If a pin is bent, do not use the 
probe until it has been inspected and repaired/replaced by 
a GE Service Representative.
•  DO NOT kink, tightly coil, or apply excessive force on the 
probe cable.  Insulation failure may result.
•  Electrical leakage checks should be performed on a routine 
basis by GE Service or qualified hospital personnel.  Refer 
to the service manual for leakage check procedures.