Getting Started
1-54 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
 Rev. 9
Probes
Introduction
Only use approved probes. 
Connecting the Probe
Probes can be connected at any time, regardless of whether the 
console is powered on or off.
To connect a probe:
1.  Place the probe's carrying case on a stable surface and 
open the case.
2.  Carefully remove the probe and unwrap the probe cord.
Inspect the probe before and after each use for damage or 
degradation to the housing, strain relief, lens, seal, cable and 
connector. DO NOT use a transducer which appears damaged 
until functional and safe performance is verified. A thorough 
inspection should be conducted during the cleaning process.
Remove any dust or foam rests from the probe pins.
Fault conditions can result in electric shock hazard. Do not 
touch the surface of probe connectors which are exposed 
when the probe is removed. Do not touch the patient when 
connecting or disconnecting a probe.
DO NOT allow the probe head to hang free. Impact to the 
probe head could result in irreparable damage.