Safety
4-22 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
 Rev. 9
Declaration of Emissions
This system is suitable for use in the following environment. The 
user must assure that it is used only in the electromagnetic 
environment as specified. 
Table 4-3:  Declaration of Emissions
Guidance and manufacturer’s declaration - electromagnetic emissions
The system is intended for use in the electromagnetic environment specified below. The 
user of the system should assure that it is used in such an environment.
Emission Type Compliance Electromagnetic Environment
RF Emissions
CISPR 11
Group 1  This system uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment. 
RF Emissions
CISPR 11
Class A This system is suitable for use in all establishments, other than 
domestic establishments and those directly connected to the 
public low-voltage power supply network that supplies buildings 
used for domestic purposes, provided the following warning is 
heeded:
WARNING: This system is intended for use by healthcare 
professionals only. This system may cause radio interference or 
may disrupt the operation of nearby equipment. It may be 
necessary to take mitigation measures, such as re-orienting or 
relocating the system or shielding the location. 
Harmonic  
Emissions
IEC 61000-3-2
Class A
Voltage 
Fluctuations/Flicker 
Emissions
IEC 61000-3-3
Complies