Measurement and Analysis
LOGIQ V2/LOGIQ V1 – User Guide 2-59
Direction 5610736-100 Rev. 9
Slope (Velocity, Time Interval and Acceleration)
To measure two velocity values, the time interval (ms), and
acceleration (m/s
2
):
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to S/D Slope.
An active caliper with a vertical and horizontal dotted line
displays.
2. To position the caliper at the start point, move the Trackball.
3. To fix the start point, press Set. The system fixes the first
caliper and displays a second active caliper.
4. To position the second caliper at the end point, move the
Trackball.
5. To complete the measurement, press Set. The system
displays the two peak end point velocities, the time interval,
and the acceleration in the Results Window.
Time interval
To measure a horizontal time interval:
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to S/D Time.
An active caliper with a vertical and horizontal dotted line
displays.
2. To position the active caliper at the start point, move the
Trackball.
3. To fix the start point, press Set. The system fixes the first
caliper and displays a second active caliper.
4. To position the second caliper at the end point, move the
Trackball.
5. To complete the measurement, press Set. The system
displays the time interval between the two calipers in the
Results Window.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals