Surgery/Intra-operative Use
LOGIQ V2/LOGIQ V1 – User Guide 5-43
Direction 5610736-100 Rev. 9
Surgery/Intra-operative Use
Preparing for Surgery/Intra-operative Procedures
Preparing the transducer for intra-operative use follows the 
same sterile procedure as for biopsy use except that no biopsy 
attachments are used. See ‘Preparing the Biopsy Guide 
Attachment’ on page 5-31 for more information. Sterile gel is 
applied to the transducer face and a sterile sheath completely 
covers the transducer and cable which has first undergone a 
thorough cleaning and high-level disinfection.
The invasive nature of biopsy procedures requires proper 
preparation and technique to control infection and disease 
transmission.  Equipment must be cleaned as appropriate for 
the procedure prior to use.
For surgery/intra-operative procedures, a sterile environment is 
required.  Therefore, both the operator and probe needs to be 
sterile.