Getting Started
1-18 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
 Rev. 9
Battery error
If there is an error on the battery, the battery error icon displays.
Figure 1-12. Battery Error Icon
Follow below steps to resolve the issue:
1.  Shutdown the system and disconnect AC power cable if it is 
connected.
2.  Remove and install the battery
3.  Connect the AC power cable
4.  Power on the system with AC power supply
5.  Disconnect AC power cable to use the battery to supply 
power to the system.
If it is still error, shutdown the system, disconnect AC power 
cable if it is connected, remove the battery and contact GE 
Service.