Probe Overview
LOGIQ V2/LOGIQ V1 – User Guide 5-13
Direction 5610736-100 Rev. 9
Disinfecting probes
Perform After 
Each Use
Ultrasound probes can be disinfected using liquid chemical 
germicides. The level of disinfection is directly related to the 
duration of contact with the germicide. Increased contact time 
produces a higher level of disinfection. Refer to the Probe Care 
Card that was shipped with each LOGIQ V2/LOGIQ V1 probe.
Ensure that you follow the probe disinfection procedure 
provided by GE.
Table 5-4:  Description of Pictogram on Probe Care Cards
Pictogram Description
“ATTENTION” - Consult accompanying documents” is intended to alert the 
user to refer to the operator manual or other instructions when complete 
information cannot be provided on the label.
“CAUTION” - Dangerous voltage (the lightning flash with arrowhead) is used to 
indicate electric shock hazards.
Biohazard - Patient/user infection due to contaminated equipment.
Usage
• Cleaning and care instructions
• Sheath and glove guidelines
Ultrasound probes are highly sensitive medical instruments that can easily be 
damaged by improper handling. Use care when handling and protect from 
damage when not in use. 
Do not immerse the probe into any liquid beyond the level specified for that 
probe. Refer to the user manual of the ultrasound system.
Since there is a possibility of having negative effects on the probe, observe the 
specified immersing time by the germicide manufacturer strictly. Do not 
immerse the probe in liquid chemical germicides more than the time prescribed 
in the care card.