Safety
4-20 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
 Rev. 9
Peripheral Update for EC countries
The following is intended to provide the users in EC countries 
with updated information concerning the connection of the 
LOGIQ V2/LOGIQ V1 to image recording and other devices or 
communication networks.
Peripherals used in the patient environment
The LOGIQ V2/LOGIQ V1 has been verified for overall safety, 
compatibility and compliance with the image recording devices 
listed in Supplies/Accessories section.
The LOGIQ V2/LOGIQ V1 has also been verified for 
compatibility, and compliance for connection to a local area 
network (LAN) via the rear panel Ethernet connection, provided 
the LAN components are IEC/EN 60950 compliant. 
The LOGIQ V2/LOGIQ V1 has also been verified for 
compatibility, and compliance for connection to a DVD-RW via 
the system USB port, provided the DVD-RW is IEC/EN 60950 
compliant.
A Wireless LAN option is available.
The LOGIQ V2/LOGIQ V1 may also be used safely while 
connected to devices other than those recommended above if 
the devices and their specifications, installation, and 
interconnection with the system conform to the requirements of 
IEC/EN 60601-1.
Accessory equipment connected to the analog and digital 
interfaces must be certified according to the respective IEC 
standards (i.e., IEC60950 for data processing equipment and 
IEC60601-1 for medical equipment). Furthermore, all complete 
configurations shall comply with the valid version of the system 
standard IEC60601-1. Everyone who connects additional 
equipment to the signal input part or signal output part of the 
LOGIQ V2/LOGIQ V1 system configures a medical system, and 
is therefore responsible to ensure that the system complies with 
the requirement of the valid version of IEC60601-1. If in doubt, 
consult the technical service department or your local GE 
representative.