Overview
LOGIQ V2/LOGIQ V1 – User Guide 1-15
Direction 5610736-100 Rev. 9
Discharge/Charge Cycle
When the battery is stored for three months or more, the
customer should perform one full discharge/charge cycle.
NOTE: A full discharge/charge cycle means the system is turned on
using battery power until the battery loses its charge completely
and the system shuts down. Plug the LOGIQ V2/LOGIQ V1 in
until the battery is fully charged as indicated by a green LCD
light.
Upon receipt of the LOGIQ V2/LOGIQ V1 and before first time
usage, it is highly recommended that the customer perform one
full discharge/charge cycle.
If the battery has not been used for >2 months, the customer is
recommended to perform one full discharge/charge cycle. It is
also recommended to store the battery in a shady and cool area
with FCC (full current capacity).
One Full Discharge/Charge Cycle Process:
1. Full discharge of battery to let the LOGIQ V2/LOGIQ V1
automatically shut down.
2. Charge the LOGIQ V2/LOGIQ V1 to 100% FCC (full current
capacity).
3. Discharge of LOGIQ V2/LOGIQ V1 for complete shut down
(takes one hour for discharge).
When storing packs for more than 6 months, charge the pack at
least once during the 6 month time frame to prevent leakage
and deterioration in performance.
CAUTION
Use only GE recognized batteries.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals