Quality Assurance
LOGIQ V2/LOGIQ V1 – User Guide 3-55
Direction 5610736-100 Rev. 9
Phantoms
Quality Assurance Evaluations may be done with phantoms and
test objects that are applicable to the parameters being
evaluated or to the user's clinical practice.
Typical phantoms are composed of material that acoustically
mimic human tissue. Pins, anechoic and echogenic targets are
physically positioned to provide information for a variety of tests.
Doppler phantoms are currently expensive and complicated to
deal with on the user level. If a problem with any Doppler
parameters or measurement is suspected, contact a local
service representative for evaluation.
The RMI 403GS phantom is shown in the illustration below as a
representative example of a phantom.