Safety
4-18 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
 Rev. 9
Notice upon Installation of Product
Separation distance and effect from fixed radio communications 
equipment: field strengths from fixed transmitters, such as base 
stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM and FM radio broadcast, and TV 
broadcast transmitter cannot be predicted theoretically with 
accuracy. To assess the electromagnetic environment due to 
fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in 
which the ultrasound system is used exceeds the applicable RF 
compliance level as stated in the immunity declaration, the 
ultrasound system should be observed to verify normal 
operation. If abnormal operation is observed, additional 
measures may be necessary, such as re-orienting or relocating 
the ultrasound system or using an RF shielded examination 
room may be necessary.
1.  Use either power supply cords provided by GE or ones 
designated by GE. Products equipped with a power source 
plug should be plugged into the fixed power socket which 
has the protective grounding conductor. Never use any 
adaptor or converter to connect with a power source plug 
(e.g. three-prong-to-two-prong converter).
2.  Locate the equipment as far away as possible from other 
electronic equipment.
3.  Be sure to use only the cables provided by or designated by 
GE. Connect these cables following the installation 
procedures (e.g. wire power cables separately from signal 
cables).
4.  Lay out the main equipment and other peripherals following 
the installation procedures described in the Option 
Installation manuals.