ActionStep
Set the desired rocking speed and angle in Settings:Rocking in the Process
Picture. Start the rocking by clicking the right-hand side of the Rocking
button.
1
Note:
When using Tray 50, the value of rocking speed multiplied by rocking angle
may not exceed 240 (e.g., with a rocking angle of 12° the rocking speed is
limited to 20 rpm).
Set the desired gas flow in Settings:Gas control:Gas flow in the Process
Picture. Start the gas flow by clicking the right-hand side of the Gas flow
button.
2
If applicable, turn on CO
2
mixing by clicking the right-hand side of the CO2
button in the Process Picture.
3
Set the required temperature setpoint.4
Click the right-hand side of the Temp button to start heating.5
Equilibrate the medium for at least 2 hours.6
140 ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
5 Operation
5.3 Prepare for cultivation
5.3.3 Add and equilibrate culture medium