DescriptionComponentPart
Power connection to the rocker.UniNet-9 port1
Indicates system connection status.CAN indicator LED2
Switch for setting the CBCU unit number for system
recognition.
CAN ID switch3
Inlet connection for O
2
supply.O2 IN4
Inlet connection for CO
2
supply.CO2 IN5
Inlet connection for air or N
2
supply.AIR/N26
CAN ID
The CAN ID is a unit number used by UNICORN to recognize the CBCU that is connected
to the system.
The CAN ID is set by turning a switch on the CBCU rear panel (see illustration above). The
CAN ID should always be set to position 1 for use in single mode. For dual mode, set the
CAN ID to 1 for the CBCU connected to the left Cellbag bioreactor, and to 2 for the CBCU
connected to the right Cellbag bioreactor.
Tubing and connectors
Tubing and connectors for gas flow as listed below are delivered with the ReadyToProcess
CBCU. Tubing and connectors for liquid flow must be obtained separately.
Tubing
LengthOuter diameterInner diameterItem
147.6" (375 cm)1/4" (6.4 mm)1/8" (3.2 mm)Tygon™ E3603
7.9" (20 cm)3/8" (9.5 mm)3/16" (4.8 mm)Silicone
Connectors
Inner diameterItem
1/8" to 3/16" (3.2 to 4.8 mm)Reducer connector, gas tubing
1/8" (3.2 mm)Connector, CBCU