Why is the power button flashing red on my GE ReadyToProcess WAVE 25 Laboratory Equipment?
- AAnita HarrisAug 22, 2025
If the power button on your GE Laboratory Equipment flashes red, it could mean several things. First, verify the connection between the rocker and the client computer or network. It's also possible that system units like the CBCU or pump defined in the system setup in UNICORN are not connected to the rocker; in this case, disable the missing components in UNICORN. Finally, an internal error may have occurred in the embedded computer of the rocker; read any warning message carefully and follow the instructions. If the problem persists, contact GE service personnel.