91
Lockout
Interval
NoNo AdjustAdjust
Note!
New Lockout Time
Requires Adjusted
Reminder Time
Adjust?
FIGURE 3-72
d) To adjust the reminder time to fit the new
QC lockout time, press Adjust. After
adjusting to an acceptable reminder time
(see FIGURE 3-76), FIGURE 3-73 will be
displayed.
To return to FIGURE 3-71, press No. The
selected time interval until the next QC
lockout will then need to be adjusted.
QC Level 1
Lockout
Next lockout
09/12/03
04:00 PM
10
11
9
12
FIGURE 3-73
The date and time of the next QC lockout
for the chosen QC level will be displayed.
e) Press the Confirm button(12) to return to
the QC Test Settings menu, FIGURE 3-67.
9 Concentration level of QC test
10 Date for next QC lockout
11 Time point for next QC lockout
12 Confirm button