2.6 The Bed Up Function Does Not Work
Chapter 2: Troubleshooting Procedures
Page 2 - 44 Century+™ Bed Service Manual (man241)
Figure 2-10. P200 Location
c. Activate the bed up switch.
d. The voltage is 12V DC (+ 3V).
Yes No
↓→Go to step 11.
7. Perform the following:
a. Disconnect the power cable to the hilow motor (B) (see figure 2-11 on
page 2-45).
m241_111