2.6 The Bed Up Function Does Not Work
Chapter 2: Troubleshooting Procedures
Century+™ Bed Service Manual (man241) Page 2 - 45
2
Figure 2-11. Hilow Motor Location
b. Set your VOM to measure volts DC.
SHOCK HAZARD:
Use extreme care when checking live voltage. Do not touch live
terminals, wires, or ground. Failure to use caution could result in
serious electrical shock injury.
c. Using the VOM, check the voltage in the power cable (see figure 3-4
on page 3-5).
d. The voltage is 12V DC +3V.
8. Check the voltage to the hilow motor. The voltage is 24V DC +3V.
Yes No
↓→Go to step 19.
9. Replace the hilow motor (B) (see figure 2-11 on page 2-45) (see “Head or
Knee Motor” on page 4-18). The bed up function now works properly.
Yes No
↓→Go to step 31.
m241_116