2.6 The Bed Up Function Does Not Work
Chapter 2: Troubleshooting Procedures
Century+™ Bed Service Manual (man241) Page 2 - 51
2
e. Check the resistance in the switch (see figure 3-3 on page 3-4, and
figure 3-4 on page 3-5).
f. While holding the VOM probes in place, depress the bed up switch.
g. The VOM reads 100Ω or less.
Yes No
↓→Replace the bed up function switch (see “Siderail Nurse Call,
Lighting, or Bed Control Switch Assembly” on page 4-52). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 31.
29. The bed up function on the bed now works properly.
Yes No
↓→Go to step 31.
30. Go to “Final Actions” on page 2-7.
31. Call Hill-Rom Technical Support at (800) 445-3720.