4.4 Resistance Measurement
────────────────────────────────────────────────────
NOTE
R
x
4
.4 Resistance Measurement
(1) Insert the measurement lead into the input terminal.
See Section 3.4.1, "Resistance Measurement", for how to connect the
measurement leads.
To measure accurately, the leads should connect correctly.
(2) Set the measurement mode to the resistance measurement.
See Section 4.2, "Selecting the Measurement Mode", for how to select the
measurement mode.
(3) Select the sampling rate and the measurement range.
(4) Execute the zero adjustment.
See Section 4.3, "Resistance Measurement Mode", for the sampling rate,
selecting the measuring range, and the zero adjustment.
(5) Connect the lead to the measured subjects (Rx), and read the measurement
value.
See Section 1.3.3, "Measurement Value Display", for measurement value
display.