4.5 Temperature Measurement
────────────────────────────────────────────────────
NOTE
4
.5 Temperature Measurement
(1) Connect the 9188 to the temperature probe connector.
See Section 3.4.2, "Temperature Correction and Temperature Measurements",
for connecting the temperature probe.
(2) Set the measurement mode to the temperature measurement.
See Section 4.2, "Selecting the Measurement Mode", for selecting the
measurement mode.
(3) Put the sensor part of the 9188 to the measuring place, and read the
measurement value.
See Section 1.3.3 "Measurement Value Display", for the display of the
measurement value.
The temperature measurement is unable to execute when the temperature
probe is unconnected or ruptured.
To prevent the inaccurate temperature measurement, do not take the
temperature probe with bare hands.
It is impossible to measure with the sensor part of the temperature probe on
the subjects. It should be used only for measuring the surrounding
temperature.
If the temperature probe is to be used in the measurement, make sure that
the 3227 and the temperature probe are thoroughly warmed up before the
measurement. If it is not used to the surrounding temperature, the serious
error is generated.