5.3 Setting the Temperature Correction
────────────────────────────────────────────────────
SHIFT
TC
SET
GP-IB ADRS
DOWN
UP
ENTER
CLEAR
ppm
Standard temperature Temperature coefficient
NOTE
5
.3 Setting the Temperature Correction
(1) Press the SHIFT key and enter the shift states. "SHIFT" on the display
lights.
(2) Press the
TC key and enter the temperature correction setting mode.
(3) Move flashing display to the setting items with the cursor keys.
(4) Set the digits of the standard temperature or the temperature coefficient
with the UP or DOWN key ("0" to "9","-").
(5) Pressing the enter key decides the setting contents and return to the
resistance measurement mode. Start executing the temperature correction
according to the setting contents.
When intend to decide the setting contents with the
ENTER key in case that
standard temperature exceeds the range of -10.0 to 99.9℃,the error is
generated. In this case, the setting contents are not decided, and the mode
does not return to the resistance measurement mode.
When the temperature conversion function is selected at power on, the setting
screen for the temperature conversion function appears.
See Section 5.5, "Setting the Temperature Conversion", for details.