64
────────────────────────────────────────────────────
7.2 Error Indications
────────────────────────────────────────────────────
The list of the error indication
Error
no.
Errors Sectio
n
01
The value exceeds 100 count is input in the zero adjustment.
Confirm the connection of the measurement lead.
4.6
02
The zero adjustment is unable to execute because using the auto range
4.6
03
The zero adjustment is unable to execute under the measurement holding state.
4.6
04
The comparator and the auto range are unable to use at the same time.
4.4, 5.
2
05
The auto range is unable to use because the external control terminal selects a
measuring range.
4.4, 6.
4
06
The temperature correction function or the temperature measurement are unable
to execute because the 9188 is not connected. Confirm the connection of the
temperature probe.
When this error occurs after connecting the temperature probe, the breaking of
wire might be generated. In this case, the temperature probe needs repair.
3.4
07
The low limit value is set bigger than the high limit value in the comparator
settings.
Confirm the settings of the high limit value and the low limit value.
1.7, 5.
1
08
The standard temperature or the initial temperature exceeds the range of 10.0
to 99.9 in the temperature correction settings or the temperature conversion
setting.
Confirm the settings of the standard temperature.
5.3, 5.
4
10
The printer is unprepared for the printer requirement.
Confirm the connection and the state of the printer.
6.5, 9.
1
7
.2 Error Indications
When there are something wrong in the self test which is performed at the
turning on the power supply, or the incorrect operations are performed for
some reason, the error indication are displayed.