EasyManua.ls Logo

Hioki 3227 - Page 77

Hioki 3227
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
65
────────────────────────────────────────────────────
7.2 Error Indications
────────────────────────────────────────────────────
Error
no.
Errors Sectio
n
11
The GP-IB address settings are unable to execute without connecting the 9588.
Connect the GP-IB interface.
3.1, 8.
1
90
Some check error of ROM is generated. It is necessary to repair this unit.
3.2
91
The data or the address error of RAM is generated.It is necessary to repair this
unit.
3.2
92
The back up error of RAM breaks the settings of this unit.
This unit executes the system reset automatically and the settings are reset to th
e
initial state.
3.2, 7.
1
93
The inaccurate measurement is generated because the proofreading data is
broken.It is necessary to repair this unit.
3.2
94
There is a fault in the internal measurement circuit. Measurements cannot be
executed. If this error persists after turning on the power supply again, the unit
requires repair.
3.2
NOTE
"Err 90", "Err 91", "Err 93, and "Err 94" show the damage occurred in the
circuit of this unit.When these errors are generated, it is necessary to repair
this unit. Please contact your nearest service representative for the repair.