67
────────────────────────────────────────────────────
8.1 Specifications
────────────────────────────────────────────────────
NOTE
Interface function
SH1
All source handshake functions
AH1
All accepter handshake functions
T6
Basic talk functions
Serial poll function
Talk-only mode is not provided.
The talker cancellation function with MLA (My Listen Address)
is provided.
L4
Basic listener function
Listen-only mode is not provided.
The listener cancellation function with MTA (My Talk Address)
is provided.
SR1
All service request functions
RL1
All remote/local functions
PP0
Parallel polling is not provided.
DC1
All device clear functions
DT1
All device trigger functions
C0
The controller function is not provided.
ASCII codes are used.
Chapter
8
GP-IB Interfac
e
8
.1 Specifications
The conform standard: IEEE-488.1 1987
The reference standard: IEEE-488.2 1987
When the output queues become full, the 9588 generates the query error and
clears them. Accordingly, this unit does not correspond to the clear of the
output queue and the query error output under the dead lock states,
stipulated in IEEE 488.2.
(The dead lock state: The state which the input buffer and the output queue
become full, and the dealing is unable to continue.)

Table of Contents