58
────────────────────────────────────────────────────
6.4 Using the Signals
────────────────────────────────────────────────────
6
.4.1 Measurement Control
T
RI
G
E
O
C
1
4msMI
N
M
easureme
nt
M
easureme
nt
F
AST: 41msMAX
M
EDIUM: 140msMA
X
S
LOW: 320msMAX
U
ntil TRIG
―――――
becomes
L
H
L
1
0
NOTE
E
O
C
M
easureme
nt
F
AST: 11msMAX
M
EDIUM: 63msMA
X
S
LOW: 250msMAX
1
0
F
AST 3msMIN
M
EDIUM, SLOW: 30msMI
N
M
easureme
nt
M
easureme
nt
6
.4 Using the Signals
(1) With measurement in the hold state
In the hold state, measurement starts when TRIG
――――――
becomes L, then EOC
becomes 1 when measurement ends. EOC then remains 1 until the next time
TRIG
――――――
becomes L.
During measurement, EOC is also invalidated when the next TRIG
――――――
signal is
input.
(2) With free-running measurement
With free-running measurement, the 3227 measures repeatedly at the internal
sampling rate, with EOC becoming 1 each time measurement ends.