25
────────────────────────────────────────────────────
3.4 Preparation for Measurements
────────────────────────────────────────────────────
3
.4.2 Temperature Correction and Temperature Measurements
NOTE
Connect the 9188 to the temperature probe connector of the rear panel with
putting the grooves of the temperature probe and that connector together,
until it clicks.
Note that the temperature probe is only designed to measure ambient air
temperature, and it is impossible to measure the surface temperature.
If the temperature probe is to be used in the measurement, make sure that
the 3227 and the temperature probe are thoroughly warmed up before the
measurement. The temperature probe and the device to be measured should
be placed close together during the warm-up, so that both will be as close to
the same temperature as possible during the measurement.