6.3 Electrical Specifications
────────────────────────────────────────────────────
BCD bit
Value
BCD bit
Value
3 2 1 0 3 2 1 0
0 0 0 0 0 0 1 0 1 5
0 0 0 1 1 0 1 1 0 6
0 0 1 0 2 0 1 1 1 7
0 0 1 1 3 1 0 0 0 8
0 1 0 0 4 1 0 0 1 9
NOTE
・ When not using the comparator or no comparator output requirement (MANU
―――――――
)
is obtained in case that the external control terminal of the comparator is
EXT, the all comparator result (Hi, IN, Lo) output "0".
・ When the digits indicated on the display are the blank (space), the digits
which are correspond to the BCD output should output "0".
・ When generated the over flow or measured under NG state, BCD output
should output "99999".
・ When no measurement has been performed after turning on the power supply,
all output signals should output "0".
・ When using the temperature conversion function, a minus sign is output as a
single BCD digit value of 1010.