146
Ref.
Code
Register Channel
Type Function Value Remark
48331 NA
Unsigned short
Bitmap for valve #49~#60:
bit[15]:valve#60 status …bit[0]:valve#49 status
0x0000 ~ 0xFFFF
New for
ACM150 G2
48332 NA
Unsigned short
Bitmap:
bit[1]: Primary pump status,
bit[0]: Secondary pump status
-
New for
ACM150 G2
48333 NA
Unsigned short
0x0001: Venturi
0x0002: Dual
0x0003: Venturi + Line leak
-
New for
ACM150 G2
48334-48335 NA
Float
(Mid-Little Endian)
Evacuation cell pressure
-
48335-48336 NA
Float
(Mid-Little Endian)
Surge cell pressure
-
ACM150 G2
48336-48337 NA
Float
(Mid-Little Endian)
Fill cell pressure
-
48401-48402 NA
Float
(Mid-Little Endian
System Running Counter (SRC)
The counter starts from 0 and is increased by
0…(2^32)-1
New for
ACM150 G2
48403-48404 NA
Float
(Mid-Little Endian
System Parameter Update Counter (SPUC)
The counter is the SRC value updated when
system parameters (ex., point enable status)
0…(2^32)-1
New for
ACM150 G2
48405-48406 NA
Float
(Mid-Little Endian
Point Parameter Update Counter (PPUC)
The counter is SRC value updated when
point#1 parameters were changed.
0…(2^32)-1
New for
ACM150 G2
48407-48408 NA
Float
(Mid-Little Endian
Point Parameter Update Counter (PPUC)
The counter is SRC value updated when
point#2 parameters were changed.
0…(2^32)-1
New for
ACM150 G2
48483-48484 NA
Float
(Mid-Little Endian
Point Parameter Update Counter (PPUC)
The counter is the SRC value updated when
point#40 parameters were changed.
0…(2^32)-1
New for
ACM150 G2
48523-48524 NA
Float
(Mid-Little Endian
Point Parameter Update Counter (PPUC)
The counter means the value of the SRC
when point#60 parameters were changed.
0…(2^32)-1
New for