4.  OPERATION
Calibration Info  Display Menu
This option displays a screen that reports information about the calibration of the Transmitter Unit 
andCertiedSensor.Anexampleofatypicalscreenisasfollows:
Calibration Info
Last Calib
368
PredictedLife: 8571
Last Calib    Displays the time in hours since the last calibration was carried out.
Next Calib     Displays the time in hours to the next calibration based on the    
  calibration interval (Calib Interval).
Gas  DisplaysthetypeofgascartridgettedtotheCertiedSensor.
FSD  Displays the maximum range of measurement.
Calib Interval    Displays the calibration interval in hours.
Predicted Life    Displays the typical life of the cartridge in hours.
CurrentCong  Display Menu
This option displays 2 screenssequentiallythatreportthecurrentTransmitterUnitandCertied
Sensorcongurationsettings,e.g.alarmsettings,relaycongurations,etc.
Initial Screen
Anexampleofatypicalinitialscreenisasfollows:
Current Config
Gas: PH3
FSD: 1.200 ppm
Alarm1
.300 ppm
Inhibit: 1.2mA
Warning: 3.0mA
Overrange 21.0mA
Gas  ReportsthetypeofgascartridgettedtotheCertiedSensor.
FSD  Displays the maximum range of gas measurement.
Alarm1  Reports the alarm value set for Alarm1.
Alarm2  Displays the alarm value set for Alarm2.
Inhibit  Reportsthevalueatwhichthe4-20mAInhibitalarmissettoactivate.
Warning  Displaysthevalueatwhichthe4-20mAWarningalarmissetto 
  activate.
Overrange  Reportsthevalueatwhichthe4-20mAOverrangealarmisset 
    to activate.