location
Filter
Sensor cap
or accessory
Lug
(3 off)
Rotate the cap/accessory in an anticlockwise direction by 1/4 turn to release the bayonet 
ttingandpulloff.
(2)  If calibrating a sensor with a non-ammable gas cartridge removethelter,iftted,fromthe
sensorcap/accessoryandtittotheFlowHousing.
Iftted,thelterisheldinplaceinthecap/accessorybyitsthreelugs.Carefullyprisethelter
free from the lug location points in the cap or accessory.
  If calibrating a sensor with a ammablegascartridgetthehydrophobicltersupplied with 
the Flow Housing into the housing.
Ineithercase,totalterintotheFlowHousing,rstremovethefactory-ttedsealfromthe
Flow Housing. The seal is held in place in the housing by its three lugs. Carefully prise the 
seal free from the lug location points in the housing.
ThelterthenlocatesintheFlowHousinginthesamewayastheremovedseal.Ensurethat
thelterisplacedintheFlowHousingthecorrectwayroundwiththethreelugsclosesttothe
front face of the housing (where the gas pipes exit) and correctly engaged in the  
location points. 
4.  OPERATION
Filter