MAN0604_Issue 09_02-2013 Apex
2110M8030
5. MAINTENANCE
Socket (2 off)
Locating
tab
recess
Sensor body
Plug
(2 off)
Locating
tab
Cartridge
(10) Pressok.
The Change Cartridge screen now displays the following message.
Change Cartridge
Working
Please Wait
This screen is displayed while the cartridge is checked by the Transmitter Unit.
When the checks are completed successfully the following type of information about the new
cartridge is displayed on the screen.
Change Cartridge
New Cartridge Type
Gas: HF
FSD: 12.00ppm
Alarm1
:1
.50ppm
Alarm2
:3
.00ppm
Checkthattheinformationdisplayedcorrespondstothetypeofreplacementcartridgetted.
(11) Press ok.
The Change Cartridge screen now displays the following message.
Change Cartridge
Success! OK:Continue
(12) Press ok.
The screen display returns to the Conguration Menu.
(13) Press esc twice.
The screen returns to the Gas Reading display.