93
MAN0604_Issue 09_02-2013 Apex
2110M8030
Notes:
1. If replacing an Oxygen cartridge with another Oxygen cartridge ensure that the Oxygen
TransducerAdaptorisleftinplaceontheCertiedSensorbody.
2. If replacing an Oxygen cartridge with a different type of cartridge ensure that the Oxygen
Transducer Adaptor is removed from the sensor body.
3. If replacing a standard cartridge with an Oxygen cartridge ensure that the Oxygen Transducer
AdaptorsuppliedwiththeOxygencartridgeisttedtotheCertiedSensorbody.Theadaptor
isttedtothesensorbodyviaabayonettting.
The Change Cartridge screen now displays the following message.
Change Cartridge
Change Cartridge Now
Cartridge: Not Present
(9) Fit the replacement cartridge.
Caution: OnlycartridgeswiththefollowingpartnumberscanbettedtotheCertied
Sensor:
2110B30x0, 31x0, 32x0, 33x0, 34x0, 35x0 series
2110B3700 - 2110B3999 range
Notes:
1. CertiedtoCSA 22.2 No. 152onlywhenttedwithspeciccartridges.
SeeSpecicationsfordetails.
CarefullyplugthecartridgeintotheCertiedSensorbodyensuringthatthecartridgetab
lines up with the recess in the sensor body and push the cartridge, without twisting, until it is
fully home.
If the cartridge will not push fully home, re-check that the locating tab is correctly aligned with
the recess in the sensor body. Position the tab so it is resting on the tab recess wall to the
right or left of the tab recess and then turn the cartridge until the tab drops into the recess.
Caution: Do not force the cartridge as this may cause damage to the pins of the connecting
plugs. Twisting and pushing can bend the pins and make the cartridge inoperative.
5. MAINTENANCE