16 810410 Field Service Bulletin XPR170
7. Use a 13/16-inch, open-ended wrench to install the fitting on the new flow meter. Tighten to
2.8 N∙m (25 in∙lbf).
Do not overtighten.
8. Hold the coolant tank fitting with a 9/16-inch, open-ended wrench.
9. Install the flow meter on the coolant tank fitting. Tighten to finger-tight, plus one turn, plus point
the larger part of the flow meter up, as shown.
10. Connect the tube to the push-to-connect fitting in the flow meter. Refer to How to use
push-to-connect fittings on page 12.
11 . Install all of the panels.
12. Add coolant to fill the tank.