5 - 10
Transpector MPH Operating Manual
5.5  Using Transpector Web UI
This section provides instruction for common Transpector MPH applications.
5.5.1  Transpector Web UI and FabGuard
FabGuard and Transpector Web UI cannot access the 
same sensor at the same time. 
FabGuard is the primary software package for the Transpector MPH sensor, but 
there are instances when using Transpector Web UI is appropriate. 
A FabGuard parameter defines when FabGuard allows Transpector Web UI to 
access the Transpector MPH sensor. This parameter has three options:
Never: Sensor Cannot Be Independently Operated
FabGuard will never give control of the Transpector MPH sensor to 
Transpector Web UI. This is the Default setting inside of FabGuard.
If Stopped: Only if the Chamber is Stopped
FabGuard will give control of the Transpector MPH sensor only if the chamber 
on which the sensor is installed is stopped.
If Stopped or Waiting: Only if the Chamber is Stopped or Waiting for 
Acquisition Trigger
FabGuard will give control of the Transpector MPH sensor if the chamber is 
stopped or if the chamber is waiting for an acquisition trigger.