7 - 1
Transpector MPH Operating Manual
Chapter 7
Diagnosing Problems
7.1  Introduction
If you are experiencing trouble with your Transpector MPH, first look at Table 7-1 
and see if your problem is listed there. If not, contact INFICON (refer to section 1.3, 
How To Contact Customer Support, on page 1-3).
7.2  Symptom-Cause-Remedy Chart
Make sure that the insulation on all cables is intact and there is no damage on the 
insulating material before using Table 7-1.
Table 7-1  Symptom—Cause—Remedy chart 
SYMPTOM CAUSE REMEDY
Power LED does not 
turn on
+24 V external power supply Check input AC line voltage 
to external power supply.
Check +24 V input, verify 
input between 20 - 30 V.
Replace power supply.
Electronics failure Return to INFICON for repair.
No communi-
cation to HOST 
computer
Sensor IP address not 
compatible with network
Refer to Chapter 2, 
Connecting 
Transpector MPH.
Transpector communications 
port (port 80) is not open on 
the host computer
Either open port 80 on the 
host computer or change the 
communications port of 
Transpector MPH.
Cable connections Make sure cables are 
connected to proper 
connectors. 
Host computer Ethernet card 
malfunctioning
Replace Ethernet Card on 
host computer.