INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 101 / 152
Type Notification Possible sources of
error
Remedy
W224 Internal voltage
24V_PRESSURE (Pirani)
voltage out of range
Short circuit or
overload in the
24V_PRESSURE
(Pirani) supply
Contact customer service
W240 Voltage +15V out of range Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
W241 Voltage -15V out of range Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
E242 +15V or -15V voltage
shorted
Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
W250 REF5V voltage out of range Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
E252 REF5V voltage shorted Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
W254 Valve current is out of range One or more lowered
valves are not
connected or defective
Contact customer service
VI-Board or MSB
defective
Contact customer service
W255 Fan current is out of range Fan not connected Contact customer service
Fan blocked or
defective
Contact customer service
W300 Anode voltage too low Short circuit of the
anode voltage
Contact customer service
VI-Board or MSB
defective
Contact customer service
W301 Anode voltage too high MSB defective Contact customer service