INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 111 / 152
Type Notification Possible sources of
error
Remedy
W710 MSB temperature is too high The ambient
temperature is too high
• Reduce the temperature in the
environment where the device is located
Air filter dirty • Clean the air vents or replace the filter
plates
Fan blocked or
defective
• Contact customer service
E711 Max. MSB temperature
exceeded
The ambient
temperature is too high
• Reduce the temperature in the
environment where the device is located
Air filter dirty • Clean the air vents or replace the filter
plates
Fan blocked or
defective
• Contact customer service
W901 Maintenance: TMP bearing/
lubricant
Maintenance interval
for TMP bearing/
lubricant exceeded
• Contact customer service
W903 Maintenance: Calibration
leak expired
Maintenance interval
for calibration leak
exceeded
• Contact customer service
W904 Maintenance: Sniffer tip filter Maintenance interval
for sniffer tip filter
exceeded
• Contact customer service
W910 Maintenance: Backing pump Maintenance interval
for backing pump
exceeded
• Contact customer service
W920 Maintenance: Exhaust filter Maintenance interval
for exhaust filter
exceeded
• Contact customer service
W925 Maintenance: Air filter Maintenance interval
for air filter exceeded
• Contact customer service