INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 111 / 152
Type Notification Possible sources of
error
Remedy
W710 MSB temperature is too high The ambient
temperature is too high
Reduce the temperature in the
environment where the device is located
Air filter dirty Clean the air vents or replace the filter
plates
Fan blocked or
defective
Contact customer service
E711 Max. MSB temperature
exceeded
The ambient
temperature is too high
Reduce the temperature in the
environment where the device is located
Air filter dirty Clean the air vents or replace the filter
plates
Fan blocked or
defective
Contact customer service
W901 Maintenance: TMP bearing/
lubricant
Maintenance interval
for TMP bearing/
lubricant exceeded
Contact customer service
W903 Maintenance: Calibration
leak expired
Maintenance interval
for calibration leak
exceeded
Contact customer service
W904 Maintenance: Sniffer tip filter Maintenance interval
for sniffer tip filter
exceeded
Contact customer service
W910 Maintenance: Backing pump Maintenance interval
for backing pump
exceeded
Contact customer service
W920 Maintenance: Exhaust filter Maintenance interval
for exhaust filter
exceeded
Contact customer service
W925 Maintenance: Air filter Maintenance interval
for air filter exceeded
Contact customer service