7 | Warning and error messages INFICON
110 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
Type Notification Possible sources of
error
Remedy
W630 Calibration request Operation mode or
mass has changed
Perform a calibration
HYDRO•S has been
switched on (if function
available)
Perform a calibration
Auto Leak Test is
active and testing time
has changed
Perform a calibration
A new SL3000 sniffer
line has been
connected (if function
available)
Perform a calibration
Runtime over 30
minutes and
temperature change by
5 Kelvin since last
calibration
Perform a calibration
Runtime over 30
minutes and no
calibration for 24 hours
Perform a calibration
W640 Auto Leak Test: Too many
consecutive tests failed
Leak rate setpoint 1
does not match the test
object
Check and change the leak rate setpoint
1 if necessary
Test chamber leaky Use another test chamber, if possible
W670 Calibration error (emission
collapsed)
During calibration the
emission collapsed
Check if the message occurs again when
you perform a new calibration
Contact customer service
W700 Preamplifier temperature is
too low
Ambient temperature is
too low
Increase the temperature in the
environment where the device is located
Temperature sensor is
defective
Contact customer service
W702 Preamplifier temperature is
too high
The ambient
temperature is too high
Reduce the temperature in the
environment where the device is located
Air filter dirty Clean the air vents or replace the filter
plates
Fan blocked or
defective
Contact customer service