INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 109 / 152
Type Notification Possible sources of
error
Remedy
W542 Flow through capillary too
high
Capillary broken or
leaky
Use another sniffer line, if possible
Contact customer service
The pressure limit for a
broken capillary is set
too low
Check and change the pressure limit, if
necessary
W559 Leak rate too high! Switched
to the standby mode to
prevent contamination!
Gross leak at test
object or the
connection to the test
object
Check the tightness of the connection
between the leak detector and the test
object
Gross leak protection
limit is set too low
Check and change the gross leak
protection limit, if necessary
Audio alarm delay time
set too short
Check and change the audio alarm delay,
if necessary
W580 Maximum evacuation time
until 100 mbar exceeded
Gross leak at test
object or the
connection to the test
object
Check the tightness of the connection
between the leak detector and the test
object
Use another test object, if possible
The settings value for
the max. gross leak
evacuation time is too
low
Check and change the max. gross leak
evacuation time if necessary
W581 Maximum evacuation time
until measurement
exceeded
Gross leak at test
object or the
connection to the test
object
Check the tightness of the connection
between the leak detector and the test
object
Use another test object, if possible
The settings value for
the max. evacuation
time until measurement
is too low
Check and change the max. evacuation
time until measurement if necessary
W625 Leak rate of internal
calibration leak not yet set
EEPROM parameter
was reset
Enter the correct leak rate for the internal
calibration leak