5 | Installation INFICON
46 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
Fig.9:
The bracket for corrugated tube is mounted.
WARNING
Shortening of the corrugated hose during evacuation
A corrugated hose connected to the inlet is shortened during evacuation.
Mount the corrugated hose in such a way that it cannot endanger the stability of the
device or flake off when it is shortened.
Mounting the exhaust hose adapter on the device
To discharge gases through a DN25 exhaust hose, mount the KF flange DN25
exhaust hose adapter supplied in the exhaust port on the rear of the device. To
seal the exhaust hose adapter, use Teflon tape. Tighten the adapter clockwise.
If you have an exhaust hose with a hose nozzle (G ½ inch male thread), you can
alternatively fix this hose nozzle in the exhaust port by turning it clockwise.