INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 97 / 152
Type Notification Possible sources of
error
Remedy
W110 Real-time clock was reset!
Enter date and time
The real-time clock has
not been set
• Enter the correct date and time
• Check that the message does not appear
when you restart the device
Battery jumper on the
MSB not set
• Contact customer service
Battery on MSB is
discharged or defective
• Contact customer service
Real-time clock
defective
• Contact customer service
W111 Many EEPROM write cycles
within the last 6 minutes
A very large number of
write commands were
issued through
interfaces. Over time,
this reduces the service
life of the EEPROM.
• Remove the excess write accesses from
your control program
E112 Impermissible DIP switch
setting
The DIP switch on the
MSB has an invalid
position
• Contact customer service
MSB defective • Contact customer service
W113 The device ID does not
match the recognized
backing pump
The DIP switch on the
MSB has an invalid
position
• Contact customer service
Type of backing pump
was not or was
incorrectly detected
• Contact customer service
MSB defective • Contact customer service
W122 No response from bus
module
Connection to BUS
module interrupted
• Check the connection to the bus module
• Replace the connection cable to the bus
module
Bus module defective • Replace the bus module
Bus module connection
on the device defective
• Contact customer service